- This
web-site is
- created
& updated
- by
Alexander Sasov
-
|
|
- SkyScan-2011:
Specification

-
-
SkyScan-2011 X-Ray NanoTomograph
SkyScan-2011
is a compact, laboratory X-Ray system for the
nondestructive three-dimensional reconstruction of the objects
internal microstructure with spatial resolution in the range of
hundreds nanometers.
PRELIMINARY SPECIFICATION
| Detail
detectability (pixel
size @ maximum magnification) |
150...250
nanometers
isotropic 3D detail detectability
(200um FOV) |
| Low-contrast resolution (10% MTF) |
400 nanometers
|
| Maximum
object size |
11mm
(8.6um pixel size) |
|
X-ray
source |
open
type (pumped) source with LaB6 cathode, 20-80kV,
100-200uA cathode current, W-target, rotary + turbomolecular pumps (typ. 2x10-7Torr), fully automatic vacuum system |
|
X-ray
detector |
Intensified
digital CCD-camera 1280x1024 pixels 12bit, FireWire
(IEEE-1394) output. |
| Computer
for system control |
Workstation with fastest Intel XEON processor(s),
2Gb RAM,
320Gb HDD (2x160GB),
20"(1600x1200) UltraSharp LCD screen |
| Reconstruction
time |
0.6sec
per cross section with 4-nodes cluster, 2sec
per cross section with single computer
(1280x1280pixels) |
|
Reconstruction method |
Volumetric
cone-beam reconstruction (Feldkamp algorithm) |
|
Radiation safety
|
<1uSv / h in any point of
the instrument surface
|
| Size
/ Weight / Operation conditions |
1300W
x 1400H x 800D mm, 540kg, 18-25C, 100-130V or 200-230V
AC, 50-60Hz, <2kW |
-
...
get application examples / page 1, JPG, 320kB >
-
...
get application examples / page 2, JPG, 270kB >
- The application examples may be seen with your
browser by scaling down to the screen size.
- To
restore image size use
button which will appear in Internet Explorer in the picture field.
-
- Alternatively you can save full format images
to your computer by using "Save
target as..." after right mouse click.
|